SMITH & NEPHEW, INC. GNS II CMT TIB SIZE 5 {} LEFT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Catalog Number 71420168
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 03/27/2013
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient was hospitalized and underwent a washout due to an infection.
 
Additional Manufacturer Narrative
It was reported that the clinical study patient was hospitalized and underwent a washout due to an infection.The affected legion oxinium femoral component, genesis ii resurfacing patellar component, journey ii articular insert and genesis ii tibial baseplate were not returned for evaluation.Therefore a product analysis could not be performed.Our investigation including a review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed parts revealed no prior complaints for the listed failure mode with the same batch number.The devices were sterilized according to sterilization release documentation from quality control.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A relationship, if any, between the devices and the reported incident or adverse event could not be corroborated.A clinical evaluation noted that the reported infection sustained less than one month after implantation was most likely due to a post-surgical complication.There is no evidence to support our device contributed to the reported infection.The patient impact beyond the washout cannot be concluded.No further clinical assessment is warranted at this time.Without the return of the actual product involved and no patient lab results available, our investigation of this report is inconclusive.No further investigation warranted for this complaint; however smith and nephew will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GNS II CMT TIB SIZE 5 {} LEFT
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8931584
Report Number1020279-2019-03197
Device Sequence Number2417493
Product Code JWH
UDI-Device Identifier03596010206275
UDI-Public03596010206275
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71420168
Device Lot Number12LT25625
Initial Date Received by Manufacturer 08/15/2019
Supplement Date Received by Manufacturer08/15/2019
Initial Report FDA Received Date08/26/2019
Supplement Report FDA Received Date11/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age63 YR
-
-