Catalog Number 329424
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Needle Stick/Puncture (2462)
Date of Event 08/07/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd micro-fine¿ insulin syringe caps were off.This occurred on 3 occasions before use.The following information was provided by the initial reporter: material no: 329424, batch no: unknown (provided 8316939).It was reported picked up the package and needles stuck her causing bleeding.(3 caps were off the sealed bag of syringes).Verbatim: patient was buying a 10 pk of needles.Picked up the package and needles stuck her causing bleeding.(3 caps were off the sealed bag of syringes).(rph) took the patient into the immunization and cleaned her up.Patient is fine.We checked the other bags inside this box and found 2 others that had some lids that come off.Soc notified.
 
Additional Manufacturer Narrative
Investigation summary: severity: s_3__; occurrence: a complaint history check was performed and this is the 1st related complaint for needle stick (before use/med intervention), harm/bleeding and cannula shield separates on lot # 8316939.Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.A review of the device history record was completed for batch# 8316939.All inspections were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.As no samples and/or photo(s) were received the investigation concluded: -unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Based on the above, no additional investigation and no capa is required at this time.
 
Event or Problem Description
It was reported that bd micro-fine¿ insulin syringe caps were off.This occurred on 3 occasions before use.The following information was provided by the initial reporter: material no: 329424 batch no: unknown (provided 8316939) it was reported picked up the package and needles stuck her causing bleeding.(3 caps were off the sealed bag of syringes).Verbatim: patient was buying a 10 pk of needles.Picked up the package and needles stuck her causing bleeding.(3 caps were off the sealed bag of syringes)._____(rph) took the patient into the immunization and cleaned her up.Patient is fine.We checked the other bags inside this box and found 2 others that had some lids that come off.-soc notified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MICRO-FINE INSULIN SYRINGE
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
MDR Report Key8931889
Report Number1920898-2019-00872
Device Sequence Number2000184
Product Code FMF
UDI-Device Identifier30382903294245
UDI-Public30382903294245
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K170386
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/31/2024
Device Catalogue Number329424
Device Lot Number8316939
Initial Date Received by Manufacturer 08/09/2019
Supplement Date Received by Manufacturer08/09/2019
Initial Report FDA Received Date08/26/2019
Supplement Report FDA Received Date09/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-