|
Returned for evaluation was a 4f dilator sheath with the guidewire still advanced inside of it.As received, the guidewire was stuck inside the 4f introducer and was contaminated with bio hazard material {dried blood} inside and out.The tip of the dilator was damaged/bent, and the guidewire appeared split.The sample was soaked in the cleaning solution.After the sample soaked, the guidewire was unable to be removed.The reported complaint is confirmed for wire stuck in introducer.The guidewire was unable to be removed from the dilator and was not fractured but split (still in one piece).A definitive root cause could not be determined due to the condition of the returned samples.A possible root cause may be the guidewire was pulled back against the needle which is cautioned against in the dfu.The dfu contains the following statements, the guidewire should not be withdrawn through the introducer needle.If the guidewire must be withdrawn while still inside the needle, simultaneously remove both the needle and the wire as a unit to prevent the needle from damaging the guidewire.A device history review of the packaging and purchased component lots revealed no quality related issues or manufacturing deficiencies at the time of manufacture.As this is considered an isolated incident, no corrective action is required.The directions for use, provided with this product, states: precautions: do not advance or withdraw a guidewire against resistance until the cause of the resistance has been determined.Excessive force against resistance may result in guidewire damage or vessel perforation.Caution: the guidewire should not be withdrawn through the introducer needle.If the guidewire must be withdrawn while still inside the needle, simultaneously remove both the needle and the wire as a unit to prevent the needle from damaging the guidewire.A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.(b)(4).
|