• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS VORTEX PORT; PORT & CATHETER, IMPLANTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS VORTEX PORT; PORT & CATHETER, IMPLANTED Back to Search Results
Catalog Number H787LVTX52130
Medical Device Problem Codes Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Hematoma (1884); Burning Sensation (2146)
Date of Event 07/24/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As no device was returned, angiodynamics is unable to perform a device evaluation.The customer's complaint description of catheter tubing detached from the port was confirmed via the chest x-ray provided.The root cause of the catheter disconnection from the port could not be determined since no sample was returned for evaluation.The end user attaches the catheter tubing and connector to the port during the implantation procedure.A review of the device history records was performed for the indicated packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly and performance specifications.The instructions for use, which is supplied to the user with this item number, contains the following statements: absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.A blood return should be present prior to usage of device for any therapy or testing.If the patient complains of pain, or if there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Caution: avoid piercing catheter with suture needle.Potential complications: catheter fragmentation and catheter pinch-off.Catheter placement considerations: warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.Pinch-off syndrome: pinch-off syndrome signs may include difficulty in aspirating blood, resistance to flushing or infusion of medications or fluids that improves with position changes, infraclavicular pain and/or swelling with catheter flushing or infusion palpitations, sudden onset chest pain, cardiac arrhythmias, extra heart sound, chest wall swelling at the port pocket, vein insertion site, pain in shoulder or port area not associated with swelling, cough, paresthesia of arm on side of catheter withdrawal occlusion or swishing sound with catheter flushing.Warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.Note: if infusion or aspiration is successful upon lifting arm above the head and turning the head, consider pinch-off syndrome as a possible cause.The line should be radiologically evaluated if pinch-off syndrome is suspected.A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint # (b)(4).
 
Event or Problem Description
As reported: dual lumen votex port inserted on (b)(6) 2019.On (b)(6) it was discovered that the catheter had become disconnected from the septum.Port was evaluated due to complaints of a large hematoma and pain by the patient.It was reported the defective disposable device is not available for return to the manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VORTEX PORT
Common Device Name
PORT & CATHETER, IMPLANTED
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
MDR Report Key8932206
Report Number1317056-2019-00109
Device Sequence Number10823654
Product Code LJT
UDI-Device IdentifierH787LVTX52130
UDI-PublicH787LVTX52130
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K010767
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician Assistant
Type of Report Initial
Report Date (Section B) 08/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberH787LVTX52130
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2019
Initial Report FDA Received Date08/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
-
-