It was reported that high impedance was detected on the patient's generator.X-rays were reviewed by the manufacturer.Per the x-rays, although the angle of the image makes it impossible to visualize the pin past the second connector block, based on the ridges in the lead pin compared to the connector blocks, the lead pin is likely fully inserted.Strain relief was not per labeling.No apparent gross lead fractures were observed; however, there was a sharp angle in the lead below the electrodes.The cause of the patient¿s high impedance could not be determined based on the images provided.Note that the presence of a micro-fracture and/or a lead could not be ruled out.Data review confirmed that high impedance had been detected on the generator.No known relevant surgical intervention has occurred to date.No further relevant surgical intervention has occurred to date.
|