| Model Number |
FPRPR3618 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/30/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The needle was returned for investigation.The manufacturing records were reviewed and no related deviation or concerns were found.The reported frost on the needle shaft was confirmed as the needle failed investigative testing.The shafts temperature is 2°c which points to insufficient thermal insulation of the needle; however the ice ball size is within the specification and was not compromised due to thermal insulation loss.Needle disassembly did not find any visible soldering gaps or voids nor corrosive signs or soldering flux residual.Based on the investigation results, the root cause determined insufficient vacuum insulation of the needle; however there was no relationship found between the needle manufacturing process and the vacuum loss phenomena.The treatment was completed with no injury to the patient.Galil medical's labeling includes precautions instructing users to protect the skin with warm saline, when appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during the freeze cycle of a cryoablation procedure, frost collected on the shaft of the needle.The patient's skin was protected and the case was completed with no injury to the patient.The needle was returned for investigation.
|
| |
|