| Catalog Number |
AC4340-01 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/24/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that, during an unspecified surgery, the tip broke off.As per reporter, no broken pieces were left inside the patient's anatomy.A back-up device was available to complete the surgery.Neither surgical delay nor patient injuries were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device, intended for use in treatment, was returned for evaluation.There was a relationship found between the device and the reported incident.Visual inspection under magnification of the wand show minimal wear on the electrode at distal end with scratch/scuff marks and discoloration on the spacer.A chunk of plastic on the spacer is missing on one side.There are no manufacturing abnormalities visually observed with the returned wand.Upon functional evaluation the device was plugged in to a good known q2 controller and was activated in a saline solution at default and maximum settings.Plasma was produced at distal end as intended.The complaint was verified and the root cause could not be determined with certainty.Factors which can contribute to the reported event: the wand is designed for ablation and/or coagulation only, and not for mechanical displacement of tissue through applied force.Do not use the wand as a lever to enlarge a surgical site or gain access to tissue as this may result in a bent or detached electrode, damage to device, and/or a cracked spacer leading to patient injury.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received, event description has been updated.
|
| |
|
Event or Problem Description
|
|
It was reported that during an ankle joint surgery, the tip broke off.As per reporter, no broken pieces were left inside the patient's anatomy.A back-up device was available to complete the surgery.Neither surgical delay nor patient injuries were reported.
|
| |
|