• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX, INC. AFX; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX, INC. AFX; BIFURCATED STENT GRAFT Back to Search Results
Model Number BA28-110/I16-30
Medical Device Problem Codes Insufficient Device Problem Information (3190); Separation Problem (4043)
Health Effect - Clinical Codes Aneurysm (1708); Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device involved in the event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx with duraply.
 
Event or Problem Description
The patient was initially implanted with a bifurcated stent graft and a suprarenal stent graft extension to treat an abdominal aortic aneurysm (aaa).Approximately 3.5 years post initial procedure (exact date was not provided), a type 3a endoleak with separation and sac growth was reported.An intervention was completed.The physician elected to bridge the main body with a suprarenal stent graft extension which successfully resolved the reported event.The patient was reported as doing fine and was discharged.
 
Additional Manufacturer Narrative
The manufacturing lot evaluation confirmed all devices met specifications prior to release.The device was not returned as the device remains implanted therefore no evaluation was completed.A clinical assessment of the reported event could not be completed due to a lack of receipt of relevant medical records and/or imaging.Several attempts were made to obtain medical records and medical images but were denied as patient authorization is needed.As such, procedure related harms, event determination, off label conditions, related patient harms and patient disposition could not be assessed.No additional investigation of this reported event is planned however if additional information pertinent to the incident is obtained, a follow-up report will be submitted.Endologix will continue to monitor this complaint and similar complaints in the event further investigation is needed.Device iteration is afx with duraply.Corrections: g1,2: contact office- name has been updated.H6: result code: remove code 3233.H6: conclusion code: remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX, INC.
2 musick
irvine CA 92618
MDR Report Key8933081
Report Number2031527-2019-00382
Device Sequence Number19450779
Product Code MIH
UDI-Device Identifier00818009014160
UDI-Public(01)00818009014160(17)181214
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 08/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/14/2018
Device Model NumberBA28-110/I16-30
Device Lot Number1432591-001
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/09/2019
Supplement Date Received by Manufacturer08/09/2019
Initial Report FDA Received Date08/26/2019
Supplement Report FDA Received Date05/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0006-2019
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VELA SUPRARENAL, LOT 1252485031
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
-
-