Model Number 304-20
Device Problems Fracture (1260); Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/01/2019
Event Type  malfunction  
Event Description
It was reported that high impedance was detected on the patient's device.The patient underwent a full revision due to high impedance.The surgeon opened the chest pocket first and found complete disintegration of lead insulation about 3 inches from the pin for about 2-3 inches.There was no indication of manipulation and no twisting of lead.The patient was in a car accident approximately a year prior and the surgeon speculated that this could be the cause.No further relevant information has been received to date.The suspect product has not been received to date.
 
Event Description
It was reported that the suspect product would not be returned.No further relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 304
Type of Device
LEAD
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
rachel kohn
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key8933214
MDR Text Key155635988
Report Number1644487-2019-01660
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750139
UDI-Public05425025750139
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/14/2018
Device Model Number304-20
Device Lot Number4111
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received08/27/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/20/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age9 YR
-
-