ABBOTT MANUFACTURING INC MAGNESIUM Back to Search Results
Catalog Number 07D70-21
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
All available patient information was included.Additional patient details are not available.An evaluation is in process.A final report will be submitted when the evaluation is complete.
 
Event or Problem Description
The customer observed a falsely elevated magnesium result on the architect c4000 analyzer.The following data was provided: initial 3.0, repeat 0.7 mmol/l.There was no impact to patient management reported.
 
Additional Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of the design history file, and a review of labeling.No adverse trend, non-conformances, potential non-conformances, or deviations were identified for the customer's issue.No return patient sample was available.Labeling was reviewed and found to be adequate.Accurate and reproducible results are dependent upon properly functioning instruments and reagents, storage of product as directed, and good laboratory technique.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNESIUM
Common Device Name
MAGNESIUM
Manufacturer (Section D)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
MDR Report Key8933332
Report Number1628664-2019-00588
Device Sequence Number2047937
Product Code JGJ
UDI-Device Identifier00380740012922
UDI-Public00380740012922
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K981791
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number07D70-21
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/13/2019
Supplement Date Received by Manufacturer10/25/2019
Initial Report FDA Received Date08/26/2019
Supplement Report FDA Received Date10/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARCHITECT C4000 ANALYZER, LN 02P24-40; ARCHITECT C4000 ANALYZER, LN 02P24-40; SERIAL # (B)(4); SERIAL # (B)(4)
-
-