| Catalog Number |
7205682 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/31/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an ankle arthroscopy the scope was bent, and no light was getting through the camera because of this there was a complete loss of visualization during the procedure.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h2, h3, h6.An evaluation was performed by the supplier and could confirm the customer complaint for the scope was bent.A visual inspection was performed and showed the scope to have a bent outertube, a broken weld and broken lenses.This damage is caused by contact with another source.No manufacturing related defects were observed.
|
| |
|