• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC PRECISION XTRA; BLOOD GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC PRECISION XTRA; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71502-01
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/19/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Unless further information is obtained, this issue is considered closed.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
 
Additional Manufacturer Narrative
An extended investigation has been conducted, which involved a manufacturing review (including 5 years of device history records (dhrs)), previous complaint and corrective and preventive action (capa) investigations conducted for blank screen issues, process failure mode effects analyses (pfmeas), design controls, and design failure mode effects analyses (dfmeas), risk management reports, risk evaluations, and label copy.The investigation did not identify any indication that the product did not meet specification.If the product is returned, an investigation will be performed.
 
Event or Problem Description
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
 
Additional Manufacturer Narrative
The returned meter was investigated with retained test strips.Meter powered on with button depression, but not test strip insertion.A port-2 issue was observed.The reported complaint was not confirmed.
 
Event or Problem Description
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION XTRA
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key8935262
Report Number2954323-2019-06595
Device Sequence Number19354881
Product Code NBW
UDI-Device Identifier00093815715023
UDI-Public(01)00093815715023(10)03W285D(17)180131
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K040814
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number71502-01
Device Catalogue Number71502
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2020
Initial Date Received by Manufacturer 08/19/2019
Supplement Date Received by Manufacturer08/30/2019
02/25/2020
Initial Report FDA Received Date08/27/2019
Supplement Report FDA Received Date09/02/2019
02/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age85 YR
Patient Weight95
-
-