| Model Number |
1403US |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/06/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.The controller was not returned for evaluation.The event file revealed controller power up event(s).As a result, the reported event was confirmed.A possible root cause of the loss of power can be attributed to a disconnection of both power sources and/or to an intermittent disconnection on both power sources.An internal investigation was initiated to capture events involving the controller losing power.This event was assessed and is being reported as part of a retrospective review of log file data.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported after log file review that the controller experienced loss of power event(s).The controller remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|