• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 1.0; VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 1.0; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 1403US
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/19/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.The controller was not returned for evaluation.The event file revealed controller power up event(s).As a result, the reported event was confirmed.A possible root cause of the loss of power can be attributed to a disconnection of both power sources and/or to an intermittent disconnection on both power sources.An internal investigation was initiated to capture events involving the controller losing power.This event was assessed and is being reported as part of a retrospective review of log file data.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported after log file review that the controller experienced loss of power event(s).The controller remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 1.0
Common Device Name
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key8937050
Report Number3007042319-2019-10432
Device Sequence Number13607449
Product Code DSQ
UDI-Device Identifier00888707000116
UDI-Public00888707000116
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/30/2017
Device Model Number1403US
Device Catalogue Number1403US
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2017
Initial Report FDA Received Date08/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-