ENCORE MEDICAL L.P. BIPOLAR HIP; SHELL, BIPOLAR, MODULAR, ENCORE, SIZE44 Back to Search Results
Model Number 412-02-044
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 07/24/2019
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to patient developing an infection, surgeon removed all implants and revised with new implants.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as an infection.The previous surgery and the surgery detailed in this event occurred 15 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history record show that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no nonconforming material reports (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was reported as an infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial).It is also possible that the patient was not compliant with post-surgical instructions.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOLAR HIP
Common Device Name
SHELL, BIPOLAR, MODULAR, ENCORE, SIZE44
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8937231
Report Number1644408-2019-00891
Device Sequence Number2876685
Product Code KWY
UDI-Device Identifier00888912077361
UDI-Public(01)00888912077361
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K953510
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/21/2021
Device Model Number412-02-044
Device Catalogue Number412-02-044
Device Lot Number988B1140
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/24/2019
Supplement Date Received by Manufacturer09/16/2019
Initial Report FDA Received Date08/27/2019
Supplement Report FDA Received Date10/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
425-96-009. LOT 188R1165; 497-28-000. LOT 629C1524
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age66 YR
-
-