• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EMPOWR 3D KNEE; EMPOWR 3D KNEETM INS, 4L 12MM, VE,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EMPOWR 3D KNEE; EMPOWR 3D KNEETM INS, 4L 12MM, VE, Back to Search Results
Model Number 341-12-704
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/29/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as loosening.The previous surgery and the revision detailed in this investigation occurred over 1 year and 7 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no nonconforming material reports (ncmr) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on- going issues that are in need of review.The root cause of this complaint was a revision surgery due to loosening.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's event.There are many factors that may contribute to the event that are outside the control of djo surgical such as poor bone density, patient bone deterioration, inadequate soft tissue support, excessive range of motion, patient activities or trauma.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - the patient was loose in flexion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR 3D KNEE
Common Device Name
EMPOWR 3D KNEETM INS, 4L 12MM, VE,
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin,tx 78758-5445
MDR Report Key8937561
Report Number1644408-2019-00798
Device Sequence Number329031
Product Code JWH
UDI-Device Identifier00888912166911
UDI-Public(01)00888912166911
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K143242
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/08/2021
Device Model Number341-12-704
Device Catalogue Number341-12-704
Device Lot Number084T1018
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2019
Initial Report FDA Received Date08/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
-
-