MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DENALI FEMORAL SYSTEM; VENA CAVA FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DENALI FEMORAL SYSTEM; VENA CAVA FILTER Back to Search Results
Model Number DL900F
Device Problems Malposition of Device (2616); Patient Device Interaction Problem (4001)
Patient Problem Death (1802)
Event Type  Death  
Manufacturer Narrative
Manufacturing review: a lot history review was performed.This is the only complaint to date for this lot number.Therefore, a device history record (dhr) review is not required.Investigation summary: the device was not returned for evaluation.Images were not provided.Medical records were provided and reviewed.Approximately four months post filter deployment, ct revealed the ivc filter was slightly tilted and the hook does not contact the ivc wall.Four of the ivc filter struts perforate the ivc anteriorly, medially, and posteriorly.As per ppf, the patient was deceased, but the cause of death and date of death was not provided in the medical record.Therefore, the investigation is confirmed for perforation of the ivc.However, the investigation is inconclusive for filter tilt as the medical records state the filter was tilted slightly and there is no objective evidence to confirm filter tilt.Based upon the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.(expiry date: 04/2018).
 
Event Description
It was reported through the litigation process that a vena cava filter was placed in a patient after being diagnosed with deep vein thrombosis/pulmonary embolism.At some time post filter deployment, it was alleged that the filter tilted and the struts perforated the ivc.The device has not been removed and there were no reported attempts made to retrieve the filter.The patient reportedly expired.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DENALI FEMORAL SYSTEM
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer Contact
judith ludwig
1625 w 3rd st.
tempe, AZ 85281
4803032689
MDR Report Key8938634
MDR Text Key155775756
Report Number2020394-2019-03000
Device Sequence Number1
Product Code DTK
UDI-Device Identifier00801741040801
UDI-Public(01)00801741040801
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDL900F
Device Catalogue NumberDL900F
Device Lot NumberGFZD0077
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/31/2019
Initial Date FDA Received08/28/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
-
-