On (b)(6) 2019, abbott point of care (apoc) was contacted by a customer regarding i-stat troponin cartridges that yielded a suspected discrepant troponin results on a (b)(6) year old male.This incident was received as part of the abbott point of care active monitoring program q3 2019.There was no patient information at the time of this report.Return product is not available for investigation.Method: i-stat, date: 03/12/2019, collected: 11:59, tested: 11:59, result: 0.00 ng/ml, sample type: wb.Lab, 03/12/2019, 11:59, 12:09, 1.7 ng/ml, ni.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.Per i-stat system manual: art: 714258-00q.Reportable range: 0.00 - 50.00.Reference range: 0.00 - 0.03 (represents the 0 to 97.5% range of results).Reference range: 0.00 - 0.08 (represents the 0 to 99% range of results).
|