• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT TROPONIN (CTNI) CARTRIDGE; CTNI CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT TROPONIN (CTNI) CARTRIDGE; CTNI CARTRIDGE Back to Search Results
Catalog Number 06P23-25
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Information (3190)
Date of Event 07/23/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.Note: this incident was received as part of the abbott point of care active monitoring program q3 2019.
 
Event or Problem Description
On (b)(6) 2019, abbott point of care (apoc) was contacted by a customer regarding i-stat troponin cartridges that yielded a suspected discrepant troponin results on a patient.This incident was received as part of the abbott point of care active monitoring program q3 2019.There was no patient information at the time of this report.Method: i-stat, date: 07/23/2019, result: 8.44.Dxi, 07/23/2019, 0.02.Collection and test time were not provided.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.Per i-stat system manual: art: 714258-00q.Reportable range: 0.00 - 50.00.Reference range: 0.00 - 0.03 (represents the 0 to 97.5% range of results).Reference range: 0.00 - 0.08 (represents the 0 to 99% range of results).
 
Additional Manufacturer Narrative
Apoc incident #: (b)(4).The investigation was completed on 09/11/2019.A review of the device history record (dhr) confirmed that the cartridge lot met finished goods (fg) release criteria.Retained cartridge testing met the acceptance outlined in appendix 1 of q04.01.003 rev.Ad (product complaint level 2 and level 3 investigation procedure).No deficiency has been identified for ctni cartridge lot g19126a.
 
Event or Problem Description
Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT TROPONIN (CTNI) CARTRIDGE
Common Device Name
CTNI CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key8939150
Report Number2245578-2019-00208
Device Sequence Number16201561
Product Code MMI
UDI-Device Identifier10054749001986
UDI-Public10054749001986
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K031739
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/05/2019
Device Catalogue Number06P23-25
Device Lot NumberG19126A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/07/2019
Supplement Date Received by Manufacturer09/11/2019
Initial Report FDA Received Date08/28/2019
Supplement Report FDA Received Date09/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-