| Catalog Number |
06P23-25 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
07/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.Note: this incident was received as part of the abbott point of care active monitoring program q3 2019.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, abbott point of care (apoc) was contacted by a customer regarding i-stat troponin cartridges that yielded a suspected discrepant troponin results on a patient.This incident was received as part of the abbott point of care active monitoring program q3 2019.There was no patient information at the time of this report.Method: i-stat, date: 07/23/2019, result: 8.44.Dxi, 07/23/2019, 0.02.Collection and test time were not provided.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.Per i-stat system manual: art: 714258-00q.Reportable range: 0.00 - 50.00.Reference range: 0.00 - 0.03 (represents the 0 to 97.5% range of results).Reference range: 0.00 - 0.08 (represents the 0 to 99% range of results).
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident #: (b)(4).The investigation was completed on 09/11/2019.A review of the device history record (dhr) confirmed that the cartridge lot met finished goods (fg) release criteria.Retained cartridge testing met the acceptance outlined in appendix 1 of q04.01.003 rev.Ad (product complaint level 2 and level 3 investigation procedure).No deficiency has been identified for ctni cartridge lot g19126a.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|