|
On 08/15/2019, abbott point of care (apoc) was contacted by a customer regarding i-stat troponin cartridges that yielded a suspected discrepant troponin results on a (b)(6) male patient.There was no additional patient information at the time of this report.Return product is not available for investigation.Method: date; collected; tested: result: positive cut-off: value: i-stat, (b)(6) 2019, 09:51, 09:51, 0.01 ng/ml, 0.08; dimension exl, (b)(6) 2019, 09:40, 09:45, 0.196 ng/ml, 0.056.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.Per i-stat system manual: art: 714258-00q reportable range: 0.00 - 50.00 reference range: 0.00 - 0.03 (represents the 0 to 97.5% range of results), reference range: 0.00 - 0.08 (represents the 0 to 99% range of results).
|