| Model Number |
G31520 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/29/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pma/510(k) #k083330.Investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
They checked that mass location was body of pancreas before the procedure in eus.They advanced echo-19 to mass in through the scope.However, the needle didn't puncture to mass in eus.Consequently, the needle removed out of the scope.So they used another echo-19.Additional information confirmed that the needle was kinked at the distal end.
|
| |
|
Additional Manufacturer Narrative
|
|
Pma/510(k)#: k083330.1 x echo-19 of lot#: c1596500 was returned to cirl for evaluation open in its original packaging.With the information provided, a physical examination and document based investigation was conducted.The device related to this occurrence underwent a laboratory evaluation.The distal tip of the needle was found to be deformed/bent.The distal end of needle was found to be kinked.The tip of sheath was found to be damaged.Likely that needle tip deformed leading to distal kink and sheathe damage (cascading effect).Prior to distribution, all echo-19 devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for echo-19 of lot number: c1596500 did not reveal any discrepancies that could have contributed to this complaint issue.The notes section of the instructions for use, which accompanies this device instructs the user to; "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".There is no evidence to suggest that the customer did not follow the instructions for use.A definitive root cause for the customer complaint could not be determined as circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to advancement into a hard lesion.Complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Event or Problem Description
|
|
They checked that mass location was body of pancreas before the procedure in eus.They advanced echo-19 to mass in through the scope.However, the needle didn't functure to mass in eus.Consequently, the needle removed out of the scope.So they used another echo-19.Additional information confirmed that the needle was kinked at the distal end.
|
| |
|