| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 21-aug-2019 the following findings: during investigation, the battery compartment was found to be cracked.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.(b)(6).
|
| |
|
Event or Problem Description
|
|
The pump was returned for investigation.Investigation revealed a battery compartment crack.This report is made based on results of investigation completed on 21-aug-2019.
|
| |
|
Search Alerts/Recalls
|