The lens was returned for evaluation.Solution is dried on the lens.The lens has been cut in half with a serrated knife, typical of insertion and removal from the eye.The lens was cleaned with lphse.A crack can be observed on each half.Multiple scratches are observed.Product history records were reviewed and the documentation indicated the product met release criteria.The root cause for the reported complaint could not be determined.The reported ¿blot¿ was not observed.It was stated the ¿blot¿ was observed when in the cartridge with a microscope; however, the lens has evidence that may indicate it was implanted and removed.Lens damage was observed.It is unknown if the damage may have been interpreted as the reported complaint.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met approved release criteria.Due to the presence of surgical solution and the condition of the returned sample, the damage observed is most likely related to customer handling.The manufacturer internal reference number is: (b)(4).
|