| Model Number |
MMT-326A |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
Hypoglycemia (1912)
|
| Date of Event |
08/19/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer called and reported that they experienced low blood glucose on (b)(6) 2019 with blood glucose of 75 mg/dl at the time of the incident.The customer used glucose tablets and food to treat.The customer did not mention any symptoms.The customer was wearing the insulin pump during the incident.The customer believes the pump was over delivering.Troubleshooting was completed but did not resolve the issue.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluated 1 random seal reservoir and checked o-rings or stopper groove for anomalies none were found.Ran basal bolus leaking test as follow: reservoir filled with diluent and connected to a new infusion set, installed reservoirs and connectors into a paradigm pump.Conclusion: reservoir not leaks and no occluded.(b)(4).
|
| |
|