| Catalog Number |
368587 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/29/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that underfill occurred during use with a bd vacutainer® precisionglide¿ multiple sample needles.The following information was provided by the initial reporter, "tube experienced under filling during use." cat# 368587.Lot# 9065849.(b)(6) 2019.(b)(6) 2019.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation: bd had not received samples, but photos were provided by the customer facility for investigation.The photos were evaluated and the customer's indicated failure mode for underfill with the incident lot was observed.Additionally, retention samples were selected from bd inventory for evaluation/testing and upon completion, the issue relating to underfill was not observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd has initiated further investigation relating to this issue through a capa#260774.The investigation is still on-going and improvements will be made as the potential causes of this issue are identified.
|
| |
|
Event or Problem Description
|
|
It was reported that underfill occurred during use with a bd vacutainer® precisionglide¿ multiple sample needles.The following information was provided by the initial reporter, "tube experienced under filling during use." cat# 368587 and lot# 9065849; 29-jul-2019, 7-aug-2019.
|
| |
|
Search Alerts/Recalls
|