| Model Number |
401750 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/08/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An orsiro drug-eluting stent system was chosen to treat a calcified lesion (>90 percent stenosis degree) in a tortuous part of the rca.The device could not be advanced.After withdrawal it was detected that both stents were damaged.A teleflex 6f guideliner was used during procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation of the returned instrument revealed that the stent is deformed at its proximal end.The balloon is well folded and shows no signs of inflation.Microscopic analysis revealed stent imprints on the exposed balloon surface indicating that the bent struts were initially crimped on the balloon.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause is most likely related to external factors during the procedure.
|
| |
|
Search Alerts/Recalls
|