| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump was returned for investigation.Investigation revealed the battery compartment had multiple cracks.This report is made based on results of investigation completed on 22-aug-2019.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.(b)(6).
|
| |
|
Event or Problem Description
|
|
The pump was returned for investigation.Investigation revealed the battery compartment was cracked.This report is made based on results of investigation completed on 22-aug-2019.There was no indication that the product caused or contributed to an adverse event.
|
| |
|