• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND PANTERA LEO 4.0/30; BASIC CORONARY ANGIOPLASTY BALLOON CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND PANTERA LEO 4.0/30; BASIC CORONARY ANGIOPLASTY BALLOON CATHETER Back to Search Results
Model Number 367043
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/08/2019
Type of Reportable Event Malfunction
Event or Problem Description
A pantera leo coronary balloon catheter was selected for treatment.During inflation the balloon could not hold the pressure.
 
Additional Manufacturer Narrative
The returned instrument was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned device revealed a fine jet of water in the distal balloon part distal to the x-ray marker when applying a pressure.The microscopic analysis of the balloon surface showed a microscopic leakage and scratches on the balloon surface nearby the damage site.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause is most likely related to the patients anatomy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PANTERA LEO 4.0/30
Common Device Name
BASIC CORONARY ANGIOPLASTY BALLOON CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key8941813
Report Number1028232-2019-03729
Device Sequence Number16774434
Product Code LOX
UDI-Device Identifier07640130416017
UDI-Public07640130416017
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K163660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2021
Device Model Number367043
Device Catalogue NumberSEE MODEL NO.
Device Lot Number04185130
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2019
Initial Date Received by Manufacturer 08/15/2019
Supplement Date Received by Manufacturer10/25/2019
Initial Report FDA Received Date08/28/2019
Supplement Report FDA Received Date10/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-