Model Number 401739
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/29/2019
Type of Reportable Event Malfunction
Event or Problem Description
An orsiro drug-eluting stent system was chosen for treatment.During preparation while removing the protection cap the stent dislodged and remained in the cap.
 
Additional Manufacturer Narrative
The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The delivery system was returned with the stent crimped on the balloon.The stent is displaced in distal direction and pinched at its distal end.Microscopic analysis of the exposed balloon surface revealed clear visible stent imprints, indicating that the stent was properly crimped in between the two radiopaque markers at the time of delivery.The balloon is well folded and shows no signs of inflation.Review of the manufacturing history of the product detailed above did not reveal any non-conformity.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.The device in question was manufactured according to specifications and successfully passed all inprocess controls as well as the final inspection.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The stent was most likely displaced on the balloon due to an improper removal of the stent protector.The ifu states that, when removing the stent protector, always pull at the very distal end of the protector to avoid dislocation of the stent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO (US) 3.5/13
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key8941823
Report Number1028232-2019-03728
Device Sequence Number11402898
Product Code NIQ
UDI-Device Identifier07640130438941
UDI-Public07640130438941
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/17/2020
Device Model Number401739
Device Catalogue NumberSEE MODEL NO.
Device Lot Number09183607
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2019
Initial Date Received by Manufacturer 08/13/2019
Supplement Date Received by Manufacturer10/17/2019
Initial Report FDA Received Date08/28/2019
Supplement Report FDA Received Date10/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-