• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ANTHOLOGY INSERTER POSTER HARD; PROSTH, HIP, SEMI-CONSTRA, UNCEMENT, METAL/POLY, NON-POROUS, CALICUM-PHOSPHATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ANTHOLOGY INSERTER POSTER HARD; PROSTH, HIP, SEMI-CONSTRA, UNCEMENT, METAL/POLY, NON-POROUS, CALICUM-PHOSPHATE Back to Search Results
Model Number 71365705
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/06/2019
Event Type  malfunction  
Event Description
It was reported that during surgery, the device broke while being used.No pieces fell into the patient's wound.Procedure was concluded with same device, no delay and no injury.
 
Manufacturer Narrative
The associated complaint device was not returned.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The strike plate on the device fractured off, rendering the device inoperable.The strike plate was not returned.The device was manufactured in 2015 and shows signs of extensive wear / usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY INSERTER POSTER HARD
Type of Device
PROSTH, HIP, SEMI-CONSTRA, UNCEMENT, METAL/POLY, NON-POROUS, CALICUM-PHOSPHATE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8942211
MDR Text Key155986112
Report Number1020279-2019-03216
Device Sequence Number1
Product Code MEH
UDI-Device Identifier03596010555663
UDI-Public03596010555663
Combination Product (y/n)N
PMA/PMN Number
K113789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 02/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71365705
Device Catalogue Number71365705
Device Lot Number15GM18622
Date Manufacturer Received01/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-