| Model Number |
MMT-523LNAP |
| Medical Device Problem Codes |
Failure to Cycle (1142); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had compromised force sensor system and drive support cap anomaly.The customer¿s blood glucose level was unknown.The customer reported that the drive support cap was sticking out.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
A test reservoir was installed and did not lock in place due to completely broken reservoir tube lip.Unable to perform rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test or verify a33 alarm due to completely broken reservoir tube lip.Device had loose or protrude drive support disk, missing reservoir tube o-ring.
|
| |
|