Model Number G31519
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) # k083330.The echo-25 device of lot number c1612330 involved in this complaint was returned for evaluation, without the original packaging.With the information provided, a physical examination and document based investigation was conducted.The device involved in the complaint was evaluated in the laboratory.The needle was found kinked, below sheath extender.Prior to distribution, all echo-25 devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl a review of the manufacturing records for echo-25 of lot number c1612330 did not reveal any discrepancies that could have contributed to this complaint issue.The notes section of the instructions for use, which accompanies this device instructs the user to; "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".There is no evidence to suggest that the customer did not follow the instructions for use.A definitive root cause could not be determined as the circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to handling/attaching the device to the scope.A capa has been initiated to document and track the actions taken to investigate and to address kinking or breaking of the sheath at the sheath/sheath extender junction complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
During use in the mediastinum lymphatic area, the needle is difficult to advance at the last 2-3 cm.The user has to forcibly apply pressure to get the needle to exit.The procedure was able to be completed successfully with this device with no patient harm.This user stated this has occurred multiple times in the last week - an additional complaint was opened to investigate this issue.Complaint device was evaluated on 31-jul-2019 and a proximal kink was observed below the sheath extender.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP ULTRA ENDOSCOPIC ULTRASOUND NEEDLE
Common Device Name
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key8944460
Report Number3001845648-2019-00443
Device Sequence Number16190076
Product Code FCG
UDI-Device Identifier00827002315198
UDI-Public(01)00827002315198(17)220510(10)C1612330
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K083330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/10/2022
Device Model NumberG31519
Device Catalogue NumberECHO-25
Device Lot NumberC1612330
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/09/2019
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2019
Initial Report FDA Received Date08/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-