| Model Number |
406844 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
During an atrial fibrillation ablation procedure a hemostasis valve leak occurred.The sheath was replaced to continue the procedure.There were no adverse patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
One 8f fast-cath introducer sheath was received for evaluation.A leak was noted at the hemostasis valve during functional testing.The cap was removed from the hemostasis hub and the hemostasis seal was microscopically inspected.Tearing, resulting in a hole, was noted on the side wall of the seal and proximal face of the seal.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn seal and subsequent leak is consistent with damage during use.The ifu states damage to the valve assembly may occur if inner catheter is withdrawn rapidly.
|
| |
|