| Model Number |
SMSBTOVL |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during laparoscopic total extraperitoneal surgery, the balloon didn't insufflate good enough.It insufflated a little bit, but not to full shape.Another device was used to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) received one device without packaging.The visual inspection of the returned product noted; the obturator, trocar balloon, lock collar and valve seal appeared intact.The insufflation bulb was received.The inflation syringe was not received.The dissector balloon was received.Pmv performed functional testing; the trocar balloon was inflated; no leaks were detected.The dissector balloon was inflated, a leak was detected.All other components functioned properly.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of hole in the dissector balloon may occur when mishandled during clinical application.The root cause of the observed damage was found to be due to the device not being used as intended which caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|