MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CR FEM LT SZ 6 NAR CEM; ATTUNE IMPLANT : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CR FEM LT SZ 6 NAR CEM; ATTUNE IMPLANT : KNEE FEMORAL Back to Search Results
Catalog Number 150400126
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Discomfort (2330); No Code Available (3191)
Date of Event 08/08/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Revision tkr; dr (b)(6) hospital (b)(6) 2019; patient underwent primary tkr approximately 3 years ago (date, surgeon and hospital unknown at present ( mater brisbane); reason for revision: pain and limited range of motion.The patient has reported that post operatively the knee did not do well and had a limited range of motion.The knee was manipulated on more than one occasion (surgeon and dates unknown), patient has presented to dr with a painful and uncomfortable knee, with a limited range of motion (+- 50 to 70 degrees).A decision was made to revise the knee (b)(6) 2019.On opening the knee the components appeared to be well fixed.The un-resurfaced patella has some unusual wear patterns.The components were removed and revised with an attune rp revision.The patella was resurfaced.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: added: evaluation codes.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FEM LT SZ 6 NAR CEM
Common Device Name
ATTUNE IMPLANT : KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key8945723
Report Number1818910-2019-102513
Device Sequence Number11272985
Product Code JWH
UDI-Device Identifier10603295041153
UDI-Public10603295041153
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K101433
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number150400126
Device Lot Number715774
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/08/2019
Supplement Date Received by Manufacturer08/29/2019
10/28/2019
Initial Report FDA Received Date08/29/2019
Supplement Report FDA Received Date09/27/2019
10/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
-
-