• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO PERFORATOR CHUCK; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO PERFORATOR CHUCK; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE Back to Search Results
Catalog Number 5400210060
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 07/18/2019
Type of Reportable Event Serious Injury
Event or Problem Description
The user facility reported three metal objects were observed in the patient's skull during a post-op scan following a procedure in which the device had been in use.The user facility reported a revision procedure was performed and 2 of the 3 metal objects were successfully retrieved.One object remains in place.There was no allegation of device failure during the initial procedure.No additional adverse consequences have been reported.
 
Additional Manufacturer Narrative
Follow-up report submitted to document device evaluation results.
 
Event or Problem Description
The user facility reported three metal objects were observed in the patient's skull during a post-op scan following a procedure in which the device had been in use.The user facility reported a revision procedure was performed and 2 of the 3 metal objects were successfully retrieved.One object remains in place.There was no allegation of device failure during the initial procedure.No additional adverse consequences have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORATOR CHUCK
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key8946471
Report Number0001811755-2019-02869
Device Sequence Number14105760
Product Code HBB
UDI-Device Identifier04546540074515
UDI-Public04546540074515
Combination Product (Y/N)N
Initial Reporter CountryGB
PMA/510(K) Number
041754
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number5400210060
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/16/2019
Supplement Date Received by Manufacturer10/22/2019
Initial Report FDA Received Date08/29/2019
Supplement Report FDA Received Date10/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-