• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS REFURB, SITE-RITE PREVUE; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS REFURB, SITE-RITE PREVUE; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770090R
Medical Device Problem Code Loss of Power (1475)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been received by the manufacturer, at this time, for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event or Problem Description
Per biomed, unit will not hold a charge.The device was charged all weekend and with 100% battery it shuts down when unplugged.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of unit will not hold a charge was confirmed, the root cause of the reported problem is a depleted internal battery.The device was serviced, tested, and returned to the customer.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event or Problem Description
Per biomed, unit will not hold a charge.The device was charged all weekend and with 100% battery it shuts down when unplugged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFURB, SITE-RITE PREVUE
Common Device Name
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8946959
Report Number3006260740-2019-02532
Device Sequence Number11377314
Product Code IYO
UDI-Device Identifier00801741124938
UDI-Public(01)00801741124938
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K150529
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 10/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9770090R
Device Catalogue Number9770090R
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2019
Event Location Hospital
Initial Date Received by Manufacturer 08/05/2019
Supplement Date Received by Manufacturer09/13/2019
Initial Report FDA Received Date08/29/2019
Supplement Report FDA Received Date10/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-