| Model Number |
9770090R |
| Medical Device Problem Code |
Loss of Power (1475)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been received by the manufacturer, at this time, for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
|
| |
|
Event or Problem Description
|
|
Per biomed, unit will not hold a charge.The device was charged all weekend and with 100% battery it shuts down when unplugged.
|
| |
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of unit will not hold a charge was confirmed, the root cause of the reported problem is a depleted internal battery.The device was serviced, tested, and returned to the customer.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
|
| |
|
Event or Problem Description
|
|
Per biomed, unit will not hold a charge.The device was charged all weekend and with 100% battery it shuts down when unplugged.
|
| |
|