Model Number 4FC12
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product event summary: the sheath 4fc12 with lot number 85396 was returned and analyzed.The visual inspection showed the device was intact with apparent issue.After cleaning and sterilizing the sheath, air aspiration was reproduced during the pressure test when the test dilator was introduced through the sheath.The hemostatic valve was leaking.Further dissection didn't show any leak along the shaft in the handle.The sheath failed the returned product inspection due to a leaking hemostatic valve.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The sheath subsequently tested out of specification per the manufacturer's investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Common Device Name
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
MDR Report Key8947031
Report Number3002648230-2019-00629
Device Sequence Number19354675
Product Code DRA
UDI-Device Identifier00763000043773
UDI-Public00763000043773
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K123591
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/22/2021
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number85396
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2019
Initial Report FDA Received Date08/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-