| Model Number |
TPS1-55-37545 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Discomfort (2330)
|
| Date of Event |
08/01/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation at this time are inconclusive as device has not been explanted.Based upon the information available and the evaluation of the x-ray images, the event appears to be due to device fatigue which could be attributed to repeated stresses on the bone screw/rod interface prior to bone fusion.It is unknown if the patient followed surgeon's post operative instructions for partial weight-bearing or use of immobilization devices.The instructions for use shares information on the increased risk of breakage of a temporary internal fixation device during postoperative rehabilitation if the patient is active.
|
| |
|
Event or Problem Description
|
|
Patient presented with a complaint of back pain.Plain film x-rays revealed a broken screw at the s1 level.Revision surgery planned to remove broken hardware and replace.
|
| |
|