Model Number MMT-523RNAH
Medical Device Problem Codes Mechanical Problem (1384); Detachment of Device or Device Component (2907); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/22/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer reported via phone call that the insulin pump had motor error alarm.The customer¿s blood glucose level was 135 mg/dl at the time of the incident.Customer stated drive support cap was sticking out.Troubleshooting was performed.Customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan per health care professional instructions.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Drive support disk was inspected, and no anomaly was noted.Device alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform rewind test, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and displacement test due to motor error alarms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key8948481
Report Number2032227-2019-55255
Device Sequence Number2860010
Product Code OYC
UDI-Device Identifier00643169221529
UDI-Public(01)00643169221529
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523RNAH
Device Catalogue NumberMMT-523RNAH
Device Lot NumberA000291727
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2019
Initial Date Received by Manufacturer 08/22/2019
Supplement Date Received by Manufacturer09/16/2019
Initial Report FDA Received Date08/30/2019
Supplement Report FDA Received Date09/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
Patient Weight220
-
-