• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522LNAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522LNAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-522LNAP
Medical Device Problem Codes Failure to Cycle (1142); Priming Problem (4040)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/22/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that she was filling her tubing when the insulin pump alarmed with compromised force sensor system alarm.The customer¿s blood glucose level was 254 mg/dl at the time of the incident.Customer stated that the drive support cap appeared normal.Customer was advised to discontinue use of the insulin pump and revert to back up plan per health care professional.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Insulin pump passed the displacement test but a33 alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform the occlusion, prime and excessive no delivery test due to a33 alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522LNAP
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key8948605
Report Number2032227-2019-55362
Device Sequence Number19496655
Product Code OYC
UDI-Device Identifier00613994407702
UDI-Public(01)00613994407702
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-522LNAP
Device Catalogue NumberMMT-522LNAP
Device Lot NumberA000114033
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2019
Initial Date Received by Manufacturer 08/23/2019
Supplement Date Received by Manufacturer09/18/2019
Initial Report FDA Received Date08/30/2019
Supplement Report FDA Received Date09/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age61 YR
Patient Weight75
-
-