| Model Number |
MMT-522LNAP |
| Medical Device Problem Codes |
Failure to Cycle (1142); Priming Problem (4040)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that she was filling her tubing when the insulin pump alarmed with compromised force sensor system alarm.The customer¿s blood glucose level was 254 mg/dl at the time of the incident.Customer stated that the drive support cap appeared normal.Customer was advised to discontinue use of the insulin pump and revert to back up plan per health care professional.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the displacement test but a33 alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform the occlusion, prime and excessive no delivery test due to a33 alarm.
|
| |
|
Search Alerts/Recalls
|