• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SRL-170
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 07/24/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Reason for reoperation: capsular contracture, baker grade iii.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
 
Event or Problem Description
Healthcare professional reported a right side capsular contracture, baker grade iii.Device remains implanted.
 
Event or Problem Description
Device was explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device.Identified: white particles.A weight test of the device was verified and the device was within specification.Bubbles after autoclave disinfection process in the gel.Based on the device analysis the final assessment is: no issues found related with the manufacturing.
 
Event or Problem Description
Healthcare professional reported a right side capsular contracture, baker grade iii.Device was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key8948888
Report Number9617229-2019-12274
Device Sequence Number2654212
Product Code FTR
UDI-Device Identifier10888628006676
UDI-Public10888628006676
Combination Product (Y/N)N
Initial Reporter StateTX
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/29/2023
Device Catalogue NumberSRL-170
Device Lot Number3207996
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2019
Initial Date Received by Manufacturer 08/02/2019
Supplement Date Received by Manufacturer09/09/2019
10/21/2019
Initial Report FDA Received Date08/30/2019
Supplement Report FDA Received Date10/07/2019
11/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age41 YR
-
-