MAUDE Adverse Event Report: INTERVASCULAR SAS HEMAGARD KNITTED; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTERVASCULAR SAS HEMAGARD KNITTED; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER Back to Search Results
Model Number HGK1407
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Stenosis (2263)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device is not accessible for testing as it remained implanted in the patient.Serial number of the graft is unknown.Therefore, no review of the device history records can be performed.The review of post-marketing historical data indicated that no other similar complaint was reported for the same lot number.The investigation is still ongoing.A follow-up report will be sent upon completion of the investigation.Not returned to manufacturer.
 
Event or Problem Description
Hemagard knitted bifurcated grafts were implanted over end of 2016 to 2017 on 21 patients.Among these patients, 7 have been recently followed up for stenosis of the graft.The date of onset of the adverse event is unknown.The stenosis area is inguinal, with almost same pattern for the 7 patients.The remaining 6 other cases are reported in the following medical device reports: 1640201-2019-00040, 1640201-2019-00041, 1640201-2019-00042, 1640201-2019-00043, 1640201-2019-00044, 1640201-2019-00045.
 
Additional Manufacturer Narrative
Corrected data: (b)(4).Serial number of the graft is unknown.Therefore, no review of the device history records can be performed¿ is replaced by: (b)(4).Serial number of the graft is unknown, therefore a complete review of the device history record cannot be performed.However the sterilization lot number associated to the product is provided.Thus, a review of the sterilization lot file associated to the device has been performed.This review concludes that the lot was released in accordance with release criteria without any nonconformities.(11) the investigation is still ongoing.A follow-up report will be sent upon completion of the investigation.
 
Event or Problem Description
See mfr report # 1640201-2019-00039.Complaint: (b)(4).
 
Event or Problem Description
See mfr report # 1640201-2019-00039 complaint: #(b)(4).The remaining 6 other cases are reported in the following medical device reports: 1640201-2019-00040, 1640201-2019-00041, 1640201-2019-00042, 1640201-2019-00043, 1640201-2019-00044, 1640201-2019-00045.
 
Additional Manufacturer Narrative
(4109/213) an additional review performed on our historical data shows no complaints for similar events in the last 2 years, except these 7 events related to 6 different lots.Therefore, it remains an isolated series of events.The clinical assessment made by our medical corporate officer is as follows : "i reviewed the information regarding seven events of stenosis in bifurcated hemagard knitted grafts that have been implanted between 2017 and 2019.In one case the stenosis led to the occlusion of the graft.In 3 cases the stenosis was not critical and did not require an intervention.In 3 cases a balloon angioplasty was completed.The events were recorded as the surgeon found unusual the occurrence of a stenosis at the same location with similar characteristics.No additional information was provided regarding pre-existing patient's conditions and surgical technique.[¿] no further comment could be made because of the lack of additional information.¿ (4315) the investigation findings do not lead to a clear conclusion about the cause of the reported adverse event.No conclusion can be drawn.However, the conducted investigation suggests that the device was not defective at the time of manufacturing.We will continue to monitor complaint trends monthly to detect any further signals.
 
Additional Manufacturer Narrative
We obtained more detailed information on the 7 patients outcome and current health condition.Among the seven patients, there were no deaths and no amputations.Also, of the seven patients, 3 had no additional procedure and 3 had angioplasty balloon.One patient who underwent angioplasty treatment recieved a stent.The last one had occlusion, therefore thrombectomy and pta were performed.Because the additionnal information provided by the hospital was not individualized, block b2 was not updated, the worst case amongst the 7 cases was considered.(4111/3221) as part of the investigation, communication/interviews with persons close to the adverse event (healthcare professional) took place in order to try to find the cause of the event.It was confirmed that the surgeon started to use hemagard in 2016.However, no more information could be furher obtained.(11) the investigation is still ongoing.A follow-up report will be sent upon completion of the investigation.
 
Event or Problem Description
See mfr report # 1640201-2019-00039.Complaint: #(b)(4).The remaining 6 other cases are reported in the following medical device reports: 1640201-2019-00040, 1640201-2019-00041, 1640201-2019-00042, 1640201-2019-00043, 1640201-2019-00044, 1640201-2019-00045.
 
Event or Problem Description
See mfr report # 1640201-2019-00039 complaint: #(b)(4).The remaining 6 other cases are reported in the following medical device reports: 1640201-2019-00040, 1640201-2019-00041, 1640201-2019-00042, 1640201-2019-00043, 1640201-2019-00044, 1640201-2019-00045.
 
Additional Manufacturer Narrative
Corrected data : block h10 : in the clinical assessment, there was a typo error confirmed today.Please read the first sentence of the clinical assessment as follows : ¿i reviewed the information regarding seven events of stenosis in bifurcated hemagard knitted grafts that have been implanted over end of 2016 to 2017.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMAGARD KNITTED
Common Device Name
PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER
Manufacturer (Section D)
INTERVASCULAR SAS
zone industrielle athelia i
la ciotat
MDR Report Key8950253
Report Number1640201-2019-00039
Device Sequence Number16762175
Product Code DSY
UDI-Device Identifier00384401014386
UDI-Public00384401014386
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K964625
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 05/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberHGK1407
Device Catalogue NumberHGK1407
Device Lot Number15G02
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/31/2019
Supplement Date Received by Manufacturer12/17/2019
02/18/2020
04/16/2020
05/11/2020
Initial Report FDA Received Date08/30/2019
Supplement Report FDA Received Date01/14/2020
03/13/2020
05/07/2020
05/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-