| Model Number |
15 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/07/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer's device and was unable to duplicate the reported issue.Physio replaced the device's battery pins as a precaution.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device unexpectedly powered off while printing.There was no report of patient use associated with the reported event.
|
| |
|