| Model Number |
AR40E |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Gender/sex: unknown, information not provided.Date of event: unknown, not provided, but the best estimate date is during 2019.If implanted, if explanted, give date: unknown, information not provided.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that intraocular lens model ar40 that were implanted had film on the edge that was visible when surgeon looked through the microscope after aspiration.Customer was not sure where the film was coming from.This report captured information for one lens.Other reports are being submitted for other lenses.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: product testing could not be performed because the product was not returned.The reported complaint was not confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.The search revealed no additional investigation request received for this production order number.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|