| Model Number |
407201 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Cardiac Perforation (2513)
|
| Date of Event |
08/22/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported cardiac perforation may have been procedure related.Per the ifu, cardiac perforation is a known risk during the use of this device.
|
| |
|
Event or Problem Description
|
|
During a pulmonary vein isolation ablation procedure, a pericardial effusion occurred.The transseptal puncture was noted to be difficult and the guidewire was seen to advance into the pericardial space on fluoroscopy and echocardiogram.Contrast was injected and moderate staining in the superior pericardium was seen.The patient remained stable and no intervention was required but the procedure was cancelled.There were no performance issues with any abbott device.
|
| |
|