ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number MF-410225
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Capsular Contracture (1761)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.The reason for reoperation: capsular contracture, baker grade iii.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
 
Event or Problem Description
Company representative reported "cap con 3".Device remains implanted.This record is for the left side.
 
Event or Problem Description
Device has been explanted.
 
Additional Manufacturer Narrative
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.Device evaluation: visual analysis of the returned device identified: weight to the spec, deformation, crease fold.Cloudy and bubbles were observed after autoclave disinfection process.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.Additional, changed, and/or corrected data: b.5, d.3, d.7, d.10, g.1, g.3, h.3, h.6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
austin TX
CS 
MDR Report Key8951148
Report Number9617229-2019-13403
Device Sequence Number2471010
Product Code FTR
UDI-Device Identifier10888628003439
UDI-Public10888628003439
Combination Product (Y/N)N
PMA/510(K) Number
P040046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/10/2020
Device Model NumberMF-410225
Device Catalogue NumberMF-410225
Device Lot Number2713610
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2020
Initial Date Received by Manufacturer 08/02/2019
Supplement Date Received by Manufacturer05/21/2020
Initial Report FDA Received Date08/30/2019
Supplement Report FDA Received Date06/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-