| Model Number |
MH693598 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
04/12/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned to the manufacturer for evaluation.The cause of the reported event cannot be determined at this time.An investigation to obtain more detailed information regarding the reported events is ongoing.As part of the investigation, a device history record (dhr) review was performed and found no abnormalities in documentation or process.The important medical information sheet provided in the device packaging, states the following, ¿this product is for single use only.It has not been designed to be re-used or re-sterilized.Reprocessing may lead to changes in material characteristics such as metallic corrosion and dulled edges, ceramic and plastic deformation, or splitting which may impact the strength of the device and compromise device performance.Reprocessing of single use devices can also cause cross contamination leading to patient infection.These risks may potentially affect patient safety.".
|
| |
|
Event or Problem Description
|
|
The company representative reported a total of 15 cases of patient infections since (b)(6) 2019 for the same facility.Two cases of infections on this 70241314 vent tube with lot # lz845980.Two cases of infections on this 70241314 vent tube with lot # nc740479.Two cases of infections on this 70241314 vent tube with lot # mh693598.Two cases of infections on this 70241314 vent tube with lot # s835440.Seven cases of infections on this 70241314 vent tube with lot # nc785867.Complaint 6 of 15.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer returned a box of 25 unopened and sealed packages of the 70241314 vent tubes (lot lz845980) for evaluation due to ¿patient infections¿.A visual inspection was performed on four of the twenty-five vent tubes that were returned, and verified the condition of the sealed, sterile package(s), which had no perforations, cuts, or irregularities.The sealing on the edge of the packages were well sealed and intact prior to opening.The condition of the blue plastic dispenser casing that encloses the actual vent tube was inspected and verified that the case opens and closes properly.Additionally, the silicone vent tube inside the case shows no signs of defects or other issues.The use by date printed on the packages are well within their time.The inner diameter and flare of the device was measured according to the information provided in the manual; the inner diameter is approximately 1.27mm and the inner flare is 1.5mm which was measure using the caliper.The cause of the reported event cannot be determined.The important medical information sheet provided in the device packaging, states the following, ¿this product is for single use only.It has not been designed to be re-used or re-sterilized.Reprocessing may lead to changes in material characteristics such as metallic corrosion and dulled edges, ceramic and plastic deformation, or splitting which may impact the strength of the device and compromise device performance.Reprocessing of single use devices can also cause cross contamination leading to patient infection.These risks may potentially affect patient safety.¿.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to report the device evaluation results and provide the device manufacture date.The customer returned 20 boxes of model 70241314.This lot was composed of two lots of blank 1186172 lot #s lz8483318 and lz843315 both of completed 600 units.These blank lots were made from silicone material 0450002 lot# me823403 which was blended on dec 5th 2018 from compound 0385012 lot #84142, blue silicone 2430090 lot # 814390 and white silicone 2430092 lot # 670703.There are no indication from receipt or dhr records that there was any abnormality with any of the materials used to create the finished good 70241314.This performance report did not find any evidence that there was any manufacturer related issue that contributed to this reported instance.Outside of going through the normal blending, molding and packaging process, the only common component between all of the reported impacted lots (nc740479, lz845980, sm835440, nc785867, mh693598) is the white silicone material 2430092 lot# 670703 was used to blend into all of the material.This material has been in use since june of 2014 with no other reported instances before this reported instance.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide additional information received, correct the event date and lot number.Please see the updates in sections: a1, b3, d4, g4, g7, h2 and h10.Additional information received from the user facility states that the patient underwent a rmt procedure.The patient continued to experience issues after the ear tubes were removed and was sent to pain management.A review of the patient¿s medical was performed and found the following: (b)(6) 2018 hd rmt ref 70241314;mh693598.(b)(6) 2019 follow up right ear otalgia (b)(6) 2019 follow up better (b)(6) 2019 follow up right ear otalgia/dizziness (b)(6) 2019 follow up right ear otalgia, muffled hearing (b)(6) 2019 surgery right ear tube removal (b)(6) 2019 follow up no changes; sent to pain management.
|
| |
|