Model Number V200
Device Problems Break (1069); No Display/Image (1183); Defective Component (2292)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2018.Date of report: 30aug2019.The field service engineer (fse) replaced the backlight inverter to address the issue.During installation of the backlight inverter, the fse noticed the pins around the keypad overlay were also damaged and replaced the keypad overlay.The unit passed extended self testing after all repairs were performed, the device passed performance verification testing.This mdr has been reassessed as reportable after a request from the fda on march 1, 2019 to review complaints from 29 nov 2017 to 29 nov 2018.As this has been reassessed, it will appear to be a late mdr.
 
Event Description
It was reported that the unit had an intermittent backlight issue that caused the display to be dim and the keypad overlay was found to be damaged causing some buttons not to work.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V200 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key8951726
MDR Text Key156450080
Report Number2031642-2019-07725
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV200
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/25/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-