• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS LEGEND TAPERED DISSECTING TOOL; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS LEGEND TAPERED DISSECTING TOOL; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number F1/8TA15
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/31/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No conclusion can be drawn.No evaluation was performed, as the devices were not returned.If the devices are returned in the future, product analysis may be performed.The user manual contains the following warning ¿do not use excessive pressure, such as bending or prying, on attachments or dissecting tools.This may cause tool to bend or break and cause injury to patient, operator and/or operating room staff.¿ we will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that two burrs broke during the procedure.It was confirmed that there was no patient impact.On follow up, it was confirmed that both tools will be returned for evaluation.
 
Additional Manufacturer Narrative
Report confirmed.Evaluation determined that two tools were returned fractured (two proximal portions and two distal portions) and used.It was noted that the fracture was relatively planar and perpendicular to the stem.A discoloration were also noted on all pieces which is most likely due to oxidation.The most likely cause was when a side load was applied/allowed while the tool was not rotating.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGEND TAPERED DISSECTING TOOL
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
MDR Report Key8952623
Report Number1625507-2019-00093
Device Sequence Number11434941
Product Code HBE
UDI-Device Identifier00643169766235
UDI-Public00643169766235
Combination Product (Y/N)N
Initial Reporter CountryLO
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/19/2024
Device Model NumberF1/8TA15
Device Catalogue NumberF1/8TA15
Device Lot Number0217070931
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2019
Supplement Date Received by Manufacturer09/23/2019
Initial Report FDA Received Date08/30/2019
Supplement Report FDA Received Date10/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-